EasyManua.ls Logo

Emerson Micro Motion 1600 - 4.7 Configure Pressure Measurement Unit; 4.7.1 Options for Pressure Measurement Unit

Emerson Micro Motion 1600
234 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
concentration measurement, Temperature Damping does not affect concentration measurement process
variables.
4.7 Configure Pressure Measurement Unit
Display MenuConfigurationProcess MeasurementPressureUnits
ProLink III
Device ToolsConfigurationProcess MeasurementPressure CompensationPressure
Unit
Field communicator Device SettingsProcess ConfigurationOptional SetupExternal Pressure/Temperature
Pressure Measurement Unit controls the measurement unit used for pressure. This unit must match the unit
used by the external pressure device.
Pressure data is used for pressure compensation and for API Referral. The device does not measure pressure
directly. You must set up a pressure input.
Procedure
Set Pressure Measurement Unit to the desired unit.
Default: psi
4.7.1 Options for Pressure Measurement Unit
The transmitter provides a standard set of measurement units for Pressure Measurement Unit. Different
communications tools may use different labels for the units. In most applications, set Pressure Measurement
Unit to match the pressure measurement unit used by the remote device.
Unit description
Label
Display ProLink III Field communicator
Feet water @ 68 °F ftH2O @68°F Ft Water @ 68°F ftH₂O
Inches water @ 4 °C inH2O @4°C In Water @ 4°C inH₂O @4DegC
Inches water @ 60 °F inH2O @60°F In Water @ 60°F inH₂O @60DegF
Inches water @ 68 °F inH2O @68°F In Water @ 68°F inH₂O
Millimeters water @ 4 °C mmH2O @4°C mm Water @ 4°C mmH₂O @4DegC
Millimeters water @ 68 °F mmH2O @68°F mm Water @ 68°F mmH₂O
Millimeters mercury @ 0 °C mmHg @0°C mm Mercury @ 0°C mmHg
Inches mercury @ 0 °C inHg @0°C In Mercury @ 0°C inHg
Pounds per square inch psi PSI psi
Bar bar bar bar
Millibar mbar millibar mbar
Grams per square centimeter g/cm2 g/cm2 g/Sqcm
Kilograms per square centimeter kg/cm2 kg/cm2 kg/Sqcm
Pascals Pa pascals Pa
Kilopascals kPA Kilopascals kPa
Megapascals mPA Megapascals MPa
Torr @ 0 °C torr Torr @ 0°C torr
Configuration and Use Manual Configure process measurement
MS-00809-0200-1600 May2024