10. Perform Smart Meter Verification.
16.6.2 Inaccurate density reading
Cause
Problem with process fluid
Incorrect density calibration factors
Wiring problem
Two-phase flow
Plugged or coated sensor tube
Incorrect sensor orientation
RTD failure
Physical characteristics of sensor have changed
Recommended actions
1. Check the wiring between the sensor and the transmitter.
2. Check the grounding of all components.
3. Check your process conditions against the values reported by the device.
4. Ensure that all of the calibration parameters have been entered correctly. See the sensor tag or the
calibration sheet for your meter.
5. Check for two-phase flow.
6. If two sensors with similar frequency are too near each other, separate them.
7. Purge the sensor tubes.
8. Perform Smart Meter Verification.
16.6.3 Unusually high density reading
Cause
Plugged or coated sensor tube
Incorrect density calibration factors
Inaccurate temperature measurement
RTD failure
In high-frequency meters, erosion, or corrosion
In low-frequency meters, tube fouling
Recommended actions
1. Ensure that all of the calibration parameters have been entered correctly.
See the sensor tag or the calibration sheet for your meter.
2. Purge the sensor tubes.
3. Check for coating in the flow tubes.
4. Perform Smart Meter Verification.
Troubleshooting Configuration and Use Manual