EasyManua.ls Logo

Emerson Micro Motion 1600 - 13.4.2 Set up pressure compensation using ProLink III

Emerson Micro Motion 1600
234 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
If you are using external temperature data, verify the external temperature value displayed in the Inputs
group on the ProLink III main window.
Choose Service ToolsVariablesExternal Variables and verify the value for External Pressure.
Choose Service ToolsVariablesProcessExternal Pressure and verify the value.
Choose Service ToolsVariablesExternal Variables and verify the value for External Temperature.
Choose Service ToolsVariablesProcessExternal Temperature and verify the value for External
Temperature.
Choose MenuService ToolsService DataView Process Variables and verify the value for External
Temperature.
Choose MenuService ToolsService DataView Process Variables and verify the values for External
Temperature and External Pressure.
Choose Service ToolsVariablesExternal Variables and verify the values for External Temperature and
External Pressure.
Choose Service ToolsVariablesProcess and verify the values for External Temperature and External
Pressure.
Need help?
If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:
Verify the wiring between the meter and the external device.
Verify the HART tag of the external device.
13.4.2 Set up pressure compensation using ProLink III
Procedure
1. Choose Device ToolsConfigurationProcess MeasurementPressure Compensation.
2. Set Pressure Compensation Status to Enabled.
3. Set Pressure Unit to the unit used by the external pressure device.
4. Enter the Density Factor and Flow Factor for your sensor.
a) Set Process Fluid to Liquid Volume or Gas Standard Volume, as appropriate.
b) Compare the values shown in Recommended Density Factor and Recommended Flow Factor to
the values from the product data sheet.
c) To use the recommended values, click Accept Recommended Values.
d) To use different factors, enter your values in the Density Factor and Flow Factor fields.
The density factor is the change in fluid density, in g/cm
3
/psi. When entering the value, reverse
the sign.
Example
If the density factor is −0.000006 g/cm
3
/psi, enter +0.000006 g/cm
3
/PSI.
The flow factor is the percent change in the flow rate per psi. When entering the value, reverse
the sign.
Measurement support Configuration and Use Manual
May2024 MS-00809-0200-1600
150 Emerson.com

Table of Contents

Other manuals for Emerson Micro Motion 1600

Related product manuals