EasyManua.ls Logo

Emerson Micro Motion 1600 - 6.2.2 Detect two-phase flow using sensor diagnostics; 6.3 Configure Flow Rate Switch

Emerson Micro Motion 1600
234 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
You must enter Two-Phase Flow Low Limit in g/cm³, even if you configured another unit for density
measurement.
Default: 0 g/cm³
Range: 0 g/cm³ to the sensor limit
2. Set Two-Phase Flow High Limit to the highest density value that is considered normal in your process.
Values above this will cause the transmitter to post a Process Aberration alert.
Tip
To reduce the occurrence of two-phase flow alerts that are not significant to your process, set Two-
Phase Flow High Limit slightly above your expected highest process density. For most applications,
leaving the Two-Phase Flow High Limit at 5.0 g/cc will suffice.
You must enter Two-Phase Flow High Limit in g/cm³, even if you configured another unit for density
measurement.
Default: 5 g/cm³
Range: 5 g/cm³ to the sensor limit
3. Set Two-Phase Flow Timeout to the number of seconds that the transmitter will wait for a two-phase
flow condition to clear before posting the alert.
Default: 0 seconds, meaning that the alert will be posted immediately
Range: 0 to 60 seconds
6.2.2 Detect two-phase flow using sensor diagnostics
Display
MenuConfigurationInputs/OutputsChannel BI/O SettingsSource
ProLink III Device ToolsConfigurationI/OChannelsChannel BmA Output
Field communicator Device SettingsInputs/OutputsMapping/SourceOutput SourcemAO x Source
The transmitter always monitors sensor diagnostics and applies a two-phase flow algorithm. You can assign
an mA Output to report the results of this calculation: single-phase flow, moderate two-phase flow, or severe
two-phase flow. Severe two-phase flow can cause the meter to stop functioning.
Prerequisites
Channel B must be configured as mA Output.
Procedure
Set mA Output Source to Two-Phase Flow Detection.
The signal from the mA Output indicates the current state of the process:
12 mA: Single-phase flow
16 mA: Moderate two-phase flow
20 mA: Severe two-phase flow
6.3 Configure Flow Rate Switch
Display
MenuConfigurationAlert SetupEnhanced EventsFlow Rate Switch
ProLink III Device ToolsConfigurationI/OOutputsDiscrete OutputSourceFlow Switch
Indication
Configure advanced options for process measurement Configuration and Use Manual
May2024 MS-00809-0200-1600
72 Emerson.com

Table of Contents

Other manuals for Emerson Micro Motion 1600

Related product manuals