Function Check Failed or Smart Meter Verification Aborted
Default severity: Maintenance Required
Severity configurable: Yes
Fault Timeout applicable: No
Table 7-9: Function Check Failed or Smart Meter Verification Aborted conditions
Name Ignorable
[010] Calibration Failed No
[034] Smart Meter Verification Failed Yes
[035] Smart Meter Verification Aborted Yes
Function Check in Progress
Default severity: Function Check
Severity configurable: No
Fault Timeout applicable: No
Table 7-10: Function Check in Progress conditions
Name Ignorable
[104] Calibration in Progress No
[131] Smart Meter Verification in Progress Yes
Output Fixed
Default severity: Function Check
Severity configurable: Yes
Fault Timeout applicable: No
Table 7-11: Output Fixed conditions
Name Ignorable
[111] Frequency Output Fixed No
[118] Discrete Output Fixed No
[101] mA Output Fixed No
Output Saturated
Default severity: Out of Speculation
Severity configurable: Yes
Fault Timeout applicable: No
Table 7-12: Output Saturated conditions
Name Ignorable
[100] mA Output Saturated Yes
[110] Frequency Output Saturated Yes
Process Aberration
Default severity: Out of Speculation
Configure device options and preferences Configuration and Use Manual