7. Perform the D2 calibration.
a) Choose D2 (Water).
b) Enter the density of your D2 fluid.
c) Choose Start Calibration.
d) Wait for the calibration to complete. This will take approximately 10 seconds.
e) Choose Finished.
8. Open the shutoff valve.
13.6.2 Perform a D1 and D2 density calibration using ProLink III
Procedure
1. Read the Prerequisites in Perform a (standard) D1 and D2 density calibration if you have not already
done so.
2. See the Figure 13-1.
Figure 13-1: D1 and D2 Calibration Work Flow Using ProLink III
D1 Calibration
Close shutoff valve
downstream from sensor
Fill sensor with D1 fluid
ProLink Menu >
Calibration >
Density Cal - Point 1
ProLink Menu >
Calibration >
Density Cal - Point 2
Enter density of D1 fluid
Do Cal Do Cal
Enter density of D2 fluid
Calibration in Progress
light turns red
Calibration in Progress
light turns green
Close
Calibration in Progress
light turns red
Calibration in Progress
light turns green
Close
Done
D2 Calibration
Fill sensor with D2 fluid
13.6.3 Perform a D1 and D2 density calibration using a field
communicator
Procedure
1. Read the Prerequisites in Perform a (standard) D1 and D2 density calibration if you have not already
done so.
Configuration and Use Manual Measurement support
MS-00809-0200-1600 May2024
Configuration and Use Manual 155