15.1.2 Smart Meter Verification history and log
The transmitter automatically saves test data for all Smart Meter Verification tests. You can generate a log
containing data for the 20 most recent tests or for all Smart Meter Verification tests. The log is an ASCII file
in .csv format.
Contents of Smart Meter Verification log
Each record in the Smart Meter Verification log represents a Smart Meter Verification test. Each record
contains the following information:
Date and time of test
Data collected during the test
The abort code (15 = test completed normally)
A pass/fail result for the left pickoff (0 = Pass, 1 = Fail)
A pass/fail result for the right pickoff (0 = Pass, 1 = Fail)
The sensor type code
The sensor serial number
Smart Meter Verification history and power-cycles
If the transmitter is rebooted or power-cycled, Smart Meter Verification history is not affected.
Smart Meter Verification history and configuration files
If you restore the factory configuration or upload a configuration file, Smart Meter Verification history is not
affected.
15.1.3 Totalizer log
The totalizer log can track four configurable totals. The period is configurable; you can configure the
transmitter to save totalizer and inventory values at a user-specified interval and then generate a totalizer
log. The totalizer log is an ASCII file.
Contents of totalizer log
The totalizer log contains one record for each logged totalizer or inventory value. Each record contains the
following information:
Default totalizer or inventory name (user-specified names are not used)
Value and measurement unit
Timestamp
Format: Military time
Time and time zone: Transmitter clock
The totalizer log also contains a line item for each totalizer or inventory reset.
Note
The historian totalizer log is configurable, but defaults to mass and volume totalizers logged at one hour
intervals.
The entire log saves 240 lines before rolling the last lines out, so 5 days of totalizer data are saved by
default.
Totalizer logs and power cycles
If the transmitter is rebooted or power-cycled, the totalizer log is not affected.
Log files, history files, and service files Configuration and Use Manual
May2024 MS-00809-0200-1600
168 Emerson.com