EasyManua.ls Logo

Emerson Micro Motion 1600 - 15.2.1 Alert history and log

Emerson Micro Motion 1600
234 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service file Description File name
Configuration Audit
Log
All changes to configuration, including changes made
by procedures such as zero calibration or density
calibration. Stores the last 1000 configuration changes.
ConfgAuditLog.txt
Alert History All occurrences of alerts and conditions, independent
of alert severity. Stores the last 1000 configuration
changes.
AlertLog.txt
Historian: 30 Days 5-minute average values for selected process and
diagnostic variables for the last 30 days.
Concatenated from
transmitter tag
and date
Historian: 1 Day Values of selected process and diagnostic variables for
the last 24 hours, recorded at 1-second intervals.
Concatenated from
transmitter tag
and date
SMV: 20 Runs Test data from the 20 most recent SMV tests. SmvLast20Data.csv
Service Snapshot An ASCII file containing a snapshot of the transmitter
internal database. This file is used by customer service.
service.dump
Factory Config File The configuration file created for this transmitter at the
factory.
FactoryConfig.cfg
Assert Log A troubleshooting file used by customer service. AssertLog.txt
Security Log A record of events that might indicate tampering. SecurityLog.txt
5. Save and specify the folder on the computer where the log file will be saved.
15.2.1 Alert history and log
The transmitter automatically saves information about all alert occurrences to its internal memory. The alert
history log is an ASCII file.
Contents of alert history
The alert history in the transmitter internal memory contains the 1000 most recent alert records. Each alert
record contains the following information:
Name of alert or condition
Category:
F = Failure
FC = Function Check
M = Maintenance Required
OOS = Out of Specification
I = Ignore
Action:
Active = Transition from inactive to active
Inactive = Transition from active to inactive
Toggling = More than 2 transitions in the last 60 seconds
Timestamp
Format: Military time
Time and time zone: Transmitter clock
Configuration and Use Manual Log files, history files, and service files