16.11.33 [131] Smart Meter Verification in Progress
Alert
Function Check in Progress
Cause
A Smart Meter Verification test is running. This can take up to 2 minutes.
Recommended actions
Allow the test to complete.
16.11.34 [132] Sensor Simulation Active
Alert
Sensor Being Simulated
Cause
Sensor simulation mode is enabled.
Recommended actions
Disable sensor simulation.
16.11.35 [138] Transient Bubble Remediation (TBR) Active
Alert
Process Aberration
Cause
Drive gain is above the configured threshold and transient bubble remediation is active.
Recommended actions
Check for air entrainment, tube fouling, flashing, or tube damage.
16.11.36 Batcher Not Configured
Alert
Configuration Error
Cause
One or more of the following has occurred:
• No flow source has been configured for the batcher application
• The batch target is 0
• No Discrete Output is assigned to batch control
Recommended actions
Verify the configuration of the batcher application and related parameters.
Troubleshooting Configuration and Use Manual
May2024 MS-00809-0200-1600