2. Exit mA Output trim, if applicable.
3. Reset HART multidrop address to zero.
4. Cycle power.
16.11.41 mA Output Saturated
Alert
Out of specification - output saturated - mA output saturated [100 or 113]
Cause
The calculated output value is outside the range of the output.
Recommended actions
1. Check the settings of Upper Range Value and Lower Range Value.
Refer to Configure Lower Range Value (LRV) and Upper Range Value (URV) for an mA Output.
2. Check your process conditions against the values reported by the device.
3. Ensure that both devices are using the same measurement unit.
4. Purge the sensor tubes.
16.11.42 Enhanced Event X Active
Alert
Out of specification - Event active - Discrete event X active
Cause
The conditions assigned to any enhanced event (Enhanced Event 1 through Enhanced Event 5) are present.
Recommended actions
1. If this is an accurate indication of process conditions, no action is required. The alert will clear when
the process returns to normal.
2. Review event configuration if you believe the event was triggered erroneously.
16.11.43 Event X Active
Alert
Out of specification - Event active - Event X active [108, 109]
Cause
The conditions assigned to either Basic Event 1 or Basic Event 2 are present.
Recommended actions
1. If this is an accurate indication of process conditions, no action is required. The alert will clear when
the process returns to normal.
2. Review event configuration if you believe the event was triggered erroneously.
16.11.44 Moderate Two Phase Flow
Alert
Process Aberration
Troubleshooting Configuration and Use Manual
May2024 MS-00809-0200-1600
204 Emerson.com