11.2.2 View and acknowledge alerts using ProLink III
You can view a list containing all alerts that are active, or inactive but unacknowledged. From this list, you can
acknowledge individual alerts or choose to acknowledge all alerts at once.
Note
Certain alerts do not clear until the transmitter is rebooted.
Procedure
1. View alerts on the ProLink III main screen under Alerts.
All active or unacknowledged alerts are listed. Take appropriate steps to clear all active alerts.
2. To acknowledge a single alert, check the Ack check box for that alert. To acknowledge all alerts at once,
select Ack All.
11.2.3 View alerts using a field communicator
You can view a list containing all alerts that are active, or inactive but unacknowledged.
Restriction
You cannot use a field communicator to acknowledge alerts. You can only view alerts. To acknowledge alerts,
use the display or make a connection to the transmitter using a different tool.
Procedure
To view active or unacknowledged alerts, choose Service ToolsAlerts.
All active alerts and unacknowledged alerts are listed. Select an alert to view detailed information.
To refresh the list, choose Service ToolsAlertsRefresh Alerts.
11.3 Read totalizer and inventory values
Display
MenuOperationsTotalizersSee Totals
ProLink III Device ToolsTotalizer ControlTotalizers
Device ToolsTotalizer ControlInventories
Field communicator Process VariablesDevice OverviewState & TasksTotalizer Control
Totalizers keep track of the total amount of mass or volume measured by the transmitter since the last
totalizer reset. Inventories keep track of the total amount of mass or volume measured by the transmitter
since the last inventory reset.
11.4 Start, stop, and reset totalizers and inventories
When a totalizer or inventory is started, its value increases or decreases depending on the interaction of the
flow direction parameters. It continues tracking flow until it is stopped.
When a totalizer or inventory is reset, its value is set to 0. You can reset a totalizer or inventory while it is
started or while it is stopped.
You can start, stop, or reset each totalizer or inventory independently.
You can start, stop, or reset all totalizers and inventories as a group.
Configuration and Use Manual Transmitter operation
MS-00809-0200-1600 May2024
Configuration and Use Manual 127