Option Description
0 Last Measured Value
1 Fault Action
3. From Register 3000, write a 6 to start DMV or a 5 to abort it.
4. Read Register 3020. Is the test running?
Option Description
> 0 = Yes Read the percent complete from Register 3020.
Range: 0 to 100
0 = No Continue to the next step.
5. Read Register 3000. Did the test run to completion?
Option
Description
0 = Yes Continue to the next step.
> 0 = No Read the abort code on Register 3002.
2 = SMV timeout — check status of pickoff and drive coils
3 = Pickoff voltage low — check status of pickoff and drive coils
4 = Temperature unstable — verify that the temperature is stable and start again
7 = Drive loop AGC reported an amplitude error — check status of pickoff and drive coils
8 = High flow (dt) standard deviation — reduce flow rate and start again
9 = High flow (dt) mean value — reduce flow rate and start again
12 = Transmitter in fault — clear alarms before proceeding
13 = No factory air verification — perform factory calibration on air
14 = No factory water verification — perform factory calibration on water
15 = Drive frequency drift from carrier frequency — ensure temperature, flow, and density
are stable and start again
6. Read Register 3004. Has the test passed?
0 = Yes
The test has passed.
> 0 = No The test has not passed. Continue to the next step.
7. Read Register 6348.
256 Possible extreme temperature or corrosion
512 Possible extreme temperature or damage
2048 Possible corrosion/erosion