EasyManuals Logo

GW Instek GPT-10000 Series User Manual

GW Instek GPT-10000 Series
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #206 background imageLoading...
Page #206 background image
GPT-10000 Series User Manual
206
SYSTem:ERRor
Query
Returns error code of the previous error. See the
error code table below for details.
SYSTem:ERRor ?
<string>
Returns an error string that includes
an error code and an error
description.
Error Code Table
Error code, Error description
0,No Error
20,Command Error
21,Value Error
22,String Error
23,Query Error
24,Mode Error
25,TIME OVER 240s
26,DC Over 50W [GPT-12XXX]
26,DC Over 100W [GPT-15XXX]
26,DC Over 120W [GPT-15012]
27,GBV > 7.2V
28,ARC <= HI Set
29,HI Set => ARC
30,Voltage Setting Error
31,Current Setting Error
32,Current HI SET Error
33,Current LO SET Error
34,Resistance HI SET Error
35,Resistance LO SET Error
36,REF Setting Error
37,Frequency Setting Error
38,ARC Setting Error
39,RAMP Time Setting Error
40,TEST Time Setting Error
41,WAIT Time Setting Error
42,RAMP Down Setting Error
43,PASS Hold Setting Error
44,GB Contact Setting Error