outside the HI SET range are ignored and will not
be used.
For example:
HI SET value: 10.99
LOW SET value: 11.00  error
MANU:DCW:CLOSet<NR2>
MANU:DCW:CLOSet?
<NR2>
0.000 ~ 10.99 (mA) [GPT-12XXX/15012]
0.000 ~ 20.99 (mA) [GPT-15XXX]
MANU:DCW:CLOS 2.00
Sets the DCW LO SET current to 2mA.
MANU:DCW:TTIMe
Set
Query
Sets or returns the DCW test time in seconds. The
test must first be in DCW mode before this
command can be used.
MANU:DCW:TTIMe {<NR2>|OFF}
MANU:DCW:TTIMe?
<NR2>
OFF
0.3 ~ 999.9 seconds
TIMER OFF
<NR2>
TIME OFF
0.3 ~ 999.9 seconds
TIMER is OFF
MANU:DCW:TTIM 1
Sets the DCW test time to 1 second.