GPT-10000 Series User Manual
218
MANU:NAME
Set
Query
Sets or returns the test name for the selected
manual test. The test must be in MANU mode
before this command can be used.
Note only alphanumeric characters (A-Z, a-z, 0-9)
and the “_” underscore character can be used to set
the MANU test name.
MANU:NAME <string>
MANU:NAME?
<string>
10 character string.
MANU:NAME “test1”
Sets the manual test name to “test1”.
MANU:RTIMe
Set
Query
Sets or returns the Ramp Up time for ACW, DCW
and IR tests in seconds.
Note: An “Err” message will be shown on display if
the Ramp Time + Test Time is > 240 seconds when
the HI SET limit + REF is ≥ 30mA for GPT-12XXX or
HI SET limit + REF is80mA for GPT-15XXX. This
applies to the ACW function only. An “TIME OVER
240s” message will be returned after using the
query command “SYSTem:ERRor ?” in remote
control.
Note: When tester is in GB or CONT mode, due to
without RAMP TIME setting, only an “Err” message
will be shown if issuing this remote command.
MANU:RTIMe <NR2>
MANU:RTIMe?
<NR2>
0.1~999.9 seconds

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-10000 Series and is the answer not in the manual?

GW Instek GPT-10000 Series Specifications

General IconGeneral
BrandGW Instek
ModelGPT-10000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals