REMOTE CONTROL
209
Example
SYST:ALERT:DATE T21_06_01
Sets the pre-alert date as 2021-06-01.
SYSTem:CAL:PROTection
Set
Query
Description
Sets or returns the calibration output protection for
calibration due date.
Syntax
Query Syntax
SYSTem:CAL:PROTection {ON |OFF}
SYSTem:CAL:PROTection?
Parameter
ON
Turns the calibration protection on.
OFF
Turns the calibration protection off.
Return parameter
ON
Calibration protection on.
OFF
Calibration protection off.
Example
SYST:CAL:PROT ON
Sets the calibration protection function ON.
SYSTem:INTernal:MEMory
Set
Query
Description
Returns the test result from internal memory of unit.
Query Syntax
SYSTem:INTernal:MEMory?
Return parameter
<string>
Returns the test result from
internal memory as a string.
Example
SYST:INTERNAL:MEMORY?
> Date/Time ,Stp,MOD,Judge, V/A , Reading ,
Timer ,MA_S,Barcode ,
2021-02-24 13:52:08,001,IR ,PASS ,0.049kV,
>10Gohm,T=000.3s,A002, ,