GPT-10000 Series User Manual
36
Test Lead Connection
This section describes how to connect the GPT-10000 to a DUT for
withstanding, insulation resistance, ground bond as well as
continuity testing.
ACW, DCW, IR Connection
Background
ACW, DCW and IR tests use the HIGH
VOLTAGE terminal and RETURN terminal
with the GHT-115 (GHT-120 for 15012) test
leads.
ACW, DCW, IR
Connection
High Voltage terminal
DUT
GPT-10000
Return terminal
Steps
1. Turn the power off on the safety analyzer.
2. Connect the high voltage test lead (red) to the
HIGH VOLTAGE terminal and screw firmly
into place.
3. Connect the return test lead (white with holder)
into the RETURN terminal and screw the
protector bar into place, as shown below.