REMOTE CONTROL
231
Sets or returns the DCW ARC function. The test
must first be in DCW mode before this command
can be used. Note that this command is only
workable when ARC SET>HI SET.
MANU:DCW:ARCFunction
{OFF|ON_CONT|ON_STOP}
MANU:DCW:ARCFunction?
Parameter/
Return parameter
MANU:DCW:ARCF OFF
Sets the DCW ARC function off.
Sets or returns the DCW ARC current value in mA.
ARC must be enabled to set the ARC current. The
test must first be in DCW mode before this
command can be used.
MANU:DCW:ARCCurrent <NR2>
MANU:DCW:ARCCurrent?
Parameter/
Return parameter
1.000 ~ 20.00 (mA) [GPT-12XXX/15012]
1.000 ~ 40.00 (mA) [GPT-15XXX]
MANU:DCW:ARCC 10
Sets the DCW ARC value to 10mA.
Sets or returns the DCW ARC speed. ARC must be
enabled before the ARC speed can be set. The test
must first be in DCW mode before this command
can be used.