Page #217 background imageLoading...
Page #217 background image
REMOTE CONTROL
217
MANU:STEP
Set
Query
Description
Sets the MANU test number.
Syntax
Query Syntax
MANU:STEP <NR1>
MANU:STEP?
Parameter/
Return parameter
<NR1>
0~100
Example
MANU:STEP 100
Sets the manual test number to 100.
MANU:INITial
Set
Query
Description
Loads the initial (default) settings for the selected
MANU test number. The initial settings that are
loaded depend on the test function (ACW, DCW,
IR, GB or CONT).
Syntax
MANU:INITial
Initial Settings
Parameter
Function
ACW
DCW
IR
GB
CONT
REF#
000uA
000uA
000.0M
Ω
000.0m
Ω
00.00Ω
FREQ
60Hz
X
X
60Hz
X
HI SET
1.000mA
1.000mA
OFF
100.0m
Ω
01.00Ω
LOW SET
000uA
000uA
000.1M
Ω
000.0m
Ω
00.00Ω
I or V
V=0.100
kV
V=0.100
kV
V=0.050
kV
03.00A
100mA
TEST TIME
000.3s
000.3s
000.3s
000.3s
000.3s
RAMP
TIME
000.1s
000.1s
000.1s
X
X
Example
MANU:INITial
Loads the initial settings for the selected MANU
number.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-10000 Series and is the answer not in the manual?

GW Instek GPT-10000 Series Specifications

General IconGeneral
BrandGW Instek
ModelGPT-10000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals