REMOTE CONTROL
241
MANU:IR:REF
Set
Query
Description
Sets or returns the IR reference value in M or GΩ.
The test must first be in IR mode before this
command can be used.
Syntax
Query Syntax
MANU:IR:REF <NR2>
MANU:IR:REF?
Parameter/
Return parameter
<NR2>
000.0M ~ 999.9M ()
1.000G ~ 9.999G ()
10.00G ~ 50.00G ()
Example
MANU:IR:REF 900M
Sets the IR reference to 900 M.
MANU:IR:MODE
Set
Query
Description
Sets or returns the IR Mode in IR. The test must
first be in IR mode before this command can be
used.
Syntax
Query Syntax
MANU:IR:MODE
{STOP_ON_FAIL|STOP_ON_PASS|TIMER}
MANU:IR:MODE?
Parameter/
Return parameter
STOP_ON_FAIL
IR Mode in Stop On FAIL
STOP_ON_PASS
IR Mode in Stop On PASS
TIMER
IR Mode in Timer
Example
MANU:IR:MODE TIMER
Sets the IR Mode in Timer setting.