OPERATION
59
Setting the Wait Time
Background
The Wait Time refers to the pending time
before FAIL judgment appears. By default,
FAIL judgment appears when test has reached
0.3 second at the earliest manner. However,
when user sets 1.0 second for Wait Time on the
tester with 0.5 second of Ramp Up time and 1.0
second of Test Time, the FAIL judgment will be
shown when Test Time has reached 0.5 second.
In short, Wait Time is the pending duration
which dominates the priority over both Ramp
UP time and Test Time in terms of timing of
FAIL judgment.
The WAIT TIME is only applicable for ACW,
DCW and IR tests.
Steps
1. Press the PAGE soft-key to move to
the 2/3 page where WAIT TIME
setting appears for ACW and DCW.
As for IR test, the WAIT TIME
setting shows in the 1/2 page.
2. Press the UP / DOWN arrow soft-
keys to bring the cursor to the
WAIT TIME setting.
WAIT TIME cursor