3. Use the scroll wheel to set the
WAIT TIME value.
ACW
DCW
IR
000.0s~999.9s
000.0s~999.9s
000.0s~999.9s
Wait Time
Indicator
While the WAIT TIME is set, the indicator of WAIT
TIME will be shown on the display in the set
duration during a test progress for clear
identification for user.
WAIT TIME indicator
Note
During the Wait Time, if test voltage is beyond the
set voltage by 1.1 times, test will be stopped
immediately in case of overvoltage issue.
During the Wait Time, if test current is beyond the
maximum currents of each unit as follows, test will be
stopped immediately in case of overcurrent issue.
GPT-12000
GPT-15000
ACW
40mA
100mA
DCW
10mA
20mA