GPT-10000 Series User Manual
222
Parameter/
Return parameter
It actually turns Off the hardware Cutoff
function and instead turns On the
software judgment function.
It remains the hardware Cutoff function On.
MANU:ACW:FILT ON
Turns On the software judgment function.
Sets or returns the ACW test time in seconds. The
test must first be in ACW mode before this
command can be used.
Note: An “Err” message will be shown on display if
the Ramp Time + Test Time is > 240 seconds when
the HI SET limit + REF is ≥ 30mA for GPT-12XXX or
HI SET limit + REF is ≥ 80mA for GPT-15XXX. This
applies to the ACW function only. An “TIME OVER
240s” message will be returned after using the
query command “SYSTem:ERRor ?” in remote
control.
MANU:ACW:TTIMe {<NR2>|OFF}
MANU:ACW:TTIMe?
0.3 ~ 999.9 seconds
TIMER OFF
0.3 ~ 999.9 seconds
TIMER is OFF
MANU:ACW:TTIM 1
Sets the ACW test time to 1 second.
Sets or returns the ACW ARC function. The test
must first be in ACW mode before this command
can be used. Note that this command is only
workable when ARC SET>HI SET.