<NR2>
000.2M ~ 999.9M ()
1.000G ~ 9.999G ()
10.00G ~ 50.00G ()
NULL
Sets the HI SET value to OFF.
MANU:IR:RHIS 10M
Sets the IR HI SET resistance to 10 M.
MANU:IR:RLOSet
Set
Query
Sets or returns the IR LO SET resistance value in
M or GΩ. The LO SET value must be less than
the HI SET value. The test must first be in IR mode
before this command can be used.
MANU:IR:RLOSet<NR2>
MANU:IR:RLOSet?
<NR1>
000.1M ~ 999.9M ()
1.000G ~ 9.999G ()
10.00G ~ 50.00G ()
MANU:IR:RLOS 10M
Sets the IR LO SET resistance to 10M.
MANU:IR:TTIMe
Set
Query
Sets or returns the IR test time in seconds. The test
must first be in IR mode before this command can
be used.
MANU:IR:TTIMe <NR2>
MANU:IR:TTIMe?
<NR2>
0.3 ~ 999.9 seconds
MANU:IR:TTIM 1
Sets the IR test time to 1 second.