EasyManuals Logo

GW Instek GPT-10000 Series User Manual

GW Instek GPT-10000 Series
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #226 background imageLoading...
Page #226 background image
GPT-10000 Series User Manual
226
MANU:ACW:MAXH OFF
Sets the ACW MAX Hold off.
MANU:ACW:PASShold
Set
Query
Sets or returns the duration of ACW PASS Hold in
seconds. The test must first be in ACW mode
before this command can be used.
MANU:ACW:PASShold {<NR2>|ON}
MANU:ACW:PASShold?
<NR2>
0 ~ 999.9 seconds
ON
Indefinite duration
MANU:ACW:PASS 999.9
Sets the ACW PASS Hold time to 999.9 seconds.