45,Setting Over 200W
46,CONT Setting Over 8V
47,Auto Step Add Full
48,This Is The Last Step
49,Learning less than 30uA
50,USB DISK BUSY
51,Filter Setting Error
70,Read Buffer Error
71,Send Buffer Error
Example
SYST:ERR ?
>0,No Error
Returns “0,No Error” as the error message.
SYSTem:CONTrol:INTerlock
Set
Query
Description
Returns the status of interlock setting.
Query Syntax
SYSTem:CONTrol:INTerlock?
Return parameter
On
Interlock setting is set On.
Off
Interlock setting is set Off.
Example
SYSTem:CONTrol:INTerlock?
> On
Returns the interlock setting is set On.
SYSTem:CAL:ALERT
Set
Query
Description
Sets or returns the calibration alert function.
Syntax
Query Syntax
SYSTem:CAL:ALERT {ON |OFF}
SYSTem:CAL:ALERT?
Parameter
ON
Turns the calibration alert on.
OFF
Turns the calibration alert off.
Return parameter
ON
Calibration alert on.
OFF
Calibration alert off.
Example
SYST:CAL:ALERT ON
Sets the calibration alert function ON.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-10000 Series and is the answer not in the manual?

GW Instek GPT-10000 Series Specifications

General IconGeneral
BrandGW Instek
ModelGPT-10000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals