REMOTE CONTROL
247
MANU:GB:REF
Set
Query
Description
Sets or returns the GB reference value in mΩ. The
test must first be in GB mode before this command
can be used.
Beware that when ISET x (HIEST + REF) is greater
than 7.2V, an “Err” message will be shown on display.
And an “GBV > 7.2V” message will be returned after
using the query command “SYSTem:ERRor ?” in
remote control.
Syntax
Query Syntax
MANU:GB:REF <NR2>
MANU:GB:REF?
Parameter/
Return parameter
<NR2>
0.000 ~ 650.0 (m)
Example
MANU:GB:REF 100
Sets the GB reference to 100 m.
MANU:GB:ZERocheck
Set
Query
Description
Performs the zero check function. The test must
first be in GB mode and in the Ready Status before
this command can be used.
See page 71 for details on the ZERO function.
Syntax
Query Syntax
MANU:GB:ZERocheck {ON|OFF}
MANU:GB:ZERocheck?
Parameter/
Return parameter
ON
Zero function is active.
OFF
Zero function is not active.
Example
MANU:GB:ZER OFF
Activates the GB ZERO function.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-10000 Series and is the answer not in the manual?

GW Instek GPT-10000 Series Specifications

General IconGeneral
BrandGW Instek
ModelGPT-10000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals