REMOTE CONTROL
235
Parameter/
Return parameter
MANU:DCW:INIT 87
Sets the DCW Initial Voltage to 87%.
Sets or returns the CONTACT CHK function on or off.
Note: An “Err” message will be shown on display if
the learned reference value is less than 30uA. An
“LEARNING < 30uA” message will be returned after
using the query command “SYSTem:ERRor ?” in
remote control.
MANU:DCW:CONTact {ON|OFF}
MANU:DCW:CONTact?
Parameter/
Return parameter
CONTACT CHK in DCW test ON
CONTACT CHK in DCW test OFF
MANU:DCW:CONT OFF
Sets the CONTACT CHK off in DCW test.
In certain applications, it is suggested to enable
Filter On, which indicates the activation of
software judgment function that judges values by
interval of every 100ms in case of instantaneous
misjudgment on distorted waveforms.
MANU:DCW:FILTer {OFF|ON}
MANU:DCW:FILTer ?