REMOTE CONTROL
225
MANU:ACW:RAMPdown
Set
Query
Description
Sets or returns the ACW Ramp Down Time in
seconds. The test must first be in ACW mode
before this command can be used.
Syntax
Query Syntax
MANU:ACW:RAMPdown <NR2>
MANU:ACW:RAMPdown?
Parameter/
Return parameter
<NR2>
0 ~ 999.9 seconds
Example
MANU:ACW:RAMP 999.9
Sets the ramp down time to 999.9 seconds.
MANU:ACW:GROundmode
Set
Query
Description
Sets or returns the ACW Ground Mode. The test
must first be in ACW mode before this command
can be used.
Syntax
Query Syntax
MANU:ACW:GROundmode {ON|OFF}
MANU:ACW:GROundmode?
Parameter/
Return parameter
ON
ACW Ground Mode ON
OFF
ACW Ground Mode OFF
Example
MANU:ACW:GRO OFF
Sets the ACW Ground Mode off.
MANU:ACW:MAXHold
Set
Query
Description
Sets or returns the ACW MAX Hold. The test must
first be in ACW mode before this command can be
used.
Syntax
Query Syntax
MANU:ACW:MAXHold {ON|OFF}
MANU:ACW:MAXHold?
Parameter/
Return parameter
ON
ACW MAX Hold ON
OFF
ACW MAX Hold OFF