EasyManuals Logo

GW Instek GPT-10000 Series User Manual

GW Instek GPT-10000 Series
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #225 background imageLoading...
Page #225 background image
REMOTE CONTROL
225
MANU:ACW:RAMPdown
Set
Query
Description
Sets or returns the ACW Ramp Down Time in
seconds. The test must first be in ACW mode
before this command can be used.
Syntax
Query Syntax
MANU:ACW:RAMPdown <NR2>
MANU:ACW:RAMPdown?
Parameter/
Return parameter
<NR2>
0 ~ 999.9 seconds
Example
MANU:ACW:RAMP 999.9
Sets the ramp down time to 999.9 seconds.
MANU:ACW:GROundmode
Set
Query
Description
Sets or returns the ACW Ground Mode. The test
must first be in ACW mode before this command
can be used.
Syntax
Query Syntax
MANU:ACW:GROundmode {ON|OFF}
MANU:ACW:GROundmode?
Parameter/
Return parameter
ON
ACW Ground Mode ON
OFF
ACW Ground Mode OFF
Example
MANU:ACW:GRO OFF
Sets the ACW Ground Mode off.
MANU:ACW:MAXHold
Set
Query
Description
Sets or returns the ACW MAX Hold. The test must
first be in ACW mode before this command can be
used.
Syntax
Query Syntax
MANU:ACW:MAXHold {ON|OFF}
MANU:ACW:MAXHold?
Parameter/
Return parameter
ON
ACW MAX Hold ON
OFF
ACW MAX Hold OFF

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-10000 Series and is the answer not in the manual?

GW Instek GPT-10000 Series Specifications

General IconGeneral
BrandGW Instek
ModelGPT-10000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals