GPT-10000 Series User Manual 
244 
MANU:GB:RHIS 100.0 
Sets the GB HI SET value to 100mΩ. 
If the GB current x (HI SET resistance + REF) > 7.2V, 
an “Err” message will be shown on display. And an 
“GBV > 7.2V” message will be returned after using the 
query command “SYSTem:ERRor ?” in remote control. 
 
Sets or returns the GB LOW SET resistance value 
in mΩ. The LOW SET value must be less than the 
HI SET value.  The test must first be in GB mode 
before this command can be used. 
MANU:GB:RLOSet<NR2> 
MANU:GB:RLOSet? 
Parameter/ 
Return parameter 
MANU:GB:RLOS 50 
Sets the GB LO SET resistance to 50mΩ. 
 
Sets or returns the GB test time in seconds. The test 
must first be in GB mode before this command can 
be used. 
MANU:GB:TTIMe <NR2> 
MANU:GB:TTIMe? 
Parameter/ 
Return parameter 
MANU:GB:TTIM 1 
Sets the GB test time to 1 second.