EasyManua.ls Logo

GW Instek GPT-10000 Series - Manu:gb:rloset; Manu:gb:ttime

GW Instek GPT-10000 Series
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GPT-10000 Series User Manual
244
MANU:GB:RHIS 100.0
Sets the GB HI SET value to 100m.
If the GB current x (HI SET resistance + REF) > 7.2V,
an “Err” message will be shown on display. And an
“GBV > 7.2V” message will be returned after using the
query command “SYSTem:ERRor ?” in remote control.
MANU:GB:RLOSet
Set
Query
Sets or returns the GB LOW SET resistance value
in m. The LOW SET value must be less than the
HI SET value. The test must first be in GB mode
before this command can be used.
MANU:GB:RLOSet<NR2>
MANU:GB:RLOSet?
<NR2>
0.000 ~ 649.9 (m)
MANU:GB:RLOS 50
Sets the GB LO SET resistance to 50m.
MANU:GB:TTIMe
Set
Query
Sets or returns the GB test time in seconds. The test
must first be in GB mode before this command can
be used.
MANU:GB:TTIMe <NR2>
MANU:GB:TTIMe?
<NR2>
0.3 ~ 999.9 seconds
MANU:GB:TTIM 1
Sets the GB test time to 1 second.

Table of Contents

Related product manuals