EasyManua.ls Logo

GW Instek GPT-10000 Series - Page 275

GW Instek GPT-10000 Series
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
APPENDIX
275
DC Withstanding Voltage
Output Voltage Range
0.050kV~ 6.000kV¹ (0.050kV~12kV for GPT-15012)
Output Voltage Resolution
1V
Output Voltage Accuracy
±(1% of setting +5V) With no load
Maximum Rated Load
GPT-15XXX
100W(5kV/20mA)
GPT-15012
120W (12kV/10mA)
GPT-12XXX
50W(5kV/10mA)
Maximum Rated Current
GPT-15XXX
20mA
0.001mA~2mA(0.05kVV0.5kV)
0.001mA~20mA(0.5kVV6kV)
GPT-15012
GPT-12XXX
10mA
0.001mA~2mA(0.05kVV0.5kV)
0.001mA~10mA(0.5kVV12kV)
10mA
0.001mA~2mA(0.05kVV0.5kV)
0.001mA~10mA(0.5kVV6kV)
Voltmeter Accuracy
± (1% of reading+ 5 V)
Voltage Regulation
± (1% +5V)[Maximum rated load ->no load]
Current Measurement Range
GPT-15XXX
0.001mA- 20.00mA
GPT-15012
GPT-12XXX
0.001mA-10.00mA
0.001mA-10.00mA
Current Measurement Best
Resolution
GPT-15XXX
GPT-15012
0.1µA
0.1µA (0.1µA~999.9µA)
1µA (1µA~9.999mA)
10µA (20.00mA)
10µA (10.00mA)
GPT-12XXX
0.1µA
0.1µA (0.1µA~999.9µA)
1µA(1µA~9.999mA)
10µA(10.00mA)
Current Measurement
Accuracy
±(1.5% of reading+3µA) when I<1mA
±(1.5% of reading+30µA) when I1mA
Judgment Accuracy
± (3% of setting+30µA)
Window Comparator Method
Yes
ARC DETECT
Yes
Rise-time Control Function
Yes
RAMP TIME (Rise Time)
0.1~999.9s
Fall-time Control Function
Yes
RAMP DOWN Time
0.0~999.9s
TIMER (Test Time)
OFF, 0.3s~999.9s
TIMER Accuracy
+/-(100ppm+20ms)
GND
ON/OFF
WAIT TIME
0.0~999.9s

Table of Contents