D-9
Flowmeter
Hardware Error 06
H Flash selftest error
Flowmeter
Hardware Error 07
H Watchdog selftest error
Flowmeter Cal.
Data Error 01
HAir, O
2
and N
2
O data empty
Flowmeter Cal.
Data Error 02
HAir, O
2
and N
2
O data error
Flowmeter Comm
Error
H XX module failed to
communicate with the main
system.
Flowmeter Comm
Stop
H The electronic flowmeter
failed to communicate with
the main system normally.
N2O Flow Too
High
LThe N
2
O flow control is
turned to set the flow too
high.
O2 Flow Too High L The O
2
flow control is turned
to set the flow too high.
Air Flow Too High L The Air flow control is turned
to set the flow too high.
Turn the flow control to
keep the flow within 10
L/min.
O2-N2O Ratio
Error
H Incorrect O2-N2O ratio. Contact your service personnel.
Flowmeter Zero
Failed
L Board or 3-way valve malfunction. Contact your service
personnel.
AG Init Error H The AG module was installed improperly or
malfunctioned.
AG Cal. Failed H The AG module calibration failed.
AG Comm Stop H AG module malfunction or communication failure
AG Airway
Occluded
H The actual pump rate of the AG module is less than 20
mL/min for more than one second.
AG Comm Error H AG module communication failure
AG Hardware Error M AG module hardware error
AG Selftest Error H Module fault or communication failure between the
module and anesthesia machine. Re-plug the module,
restart the anesthesia machine, or try to plug the module
into another anesthesia machine.
AG
module
AG Hardware
Malfunction
H AG module hardware malfunction. The AG module enters
Standby and measurement stops. Remove the AG module
and contact your service personnel.