D-10
AG Watertrap Type
Wrong
M The watertrap of the AG module is of wrong type.
Replace with a correct watertrap.
AG Data Limit
Error
M AG module malfunction
AG Accuracy Error M The measured value is outside of the measurement
accuracy range.
FiO2 ALM LMT
ERR
M The FiO2 alarm limit settings are outside of the range.
EtCO2 ALM LMT
ERR
MThe EtCO
2
alarm limit settings are outside of the range.
FiCO2 ALM LMT
ERR
M The FiCO2 alarm limit settings are outside of the range.
EtN2O ALM LMT
ERR
MThe EtN
2
O alarm limit settings are outside of the range.
FiN2O ALM LMT
ERR
MThe FiN
2
O alarm limit settings are outside of the range.
EtAA ALM LMT
ERR
M The EtAA alarm limit settings are outside of the range.
FiAA ALM LMT
ERR
M The FiAA alarm limit settings are outside of the range.
AG No Watertrap L The AG watertrap fell off from the anesthesia machine.
AG Zero Failed L The AG module zeroing failed.
AG Change
Watertrap
M The AG watertrap was changed.
EtCO2 Overrange H
FiCO2 Overrange H
EtN2O Overrange H
FiN2O Overrange H
EtAA Overrange H
FiAA Overrange H
The measured value is outside of the measurement range.
Contact your service personnel.
CO2 Accuracy
Unspecified
L
O2 Accuracy
Unspecified
L
N2O Accuracy
Unspecified
L
AA Accuracy
Unspecified
L
The measured value is outside of the declared accuracy
range.
Mixed Agent and L More than more anesthetic agents were detected ant the