D-12 
CO2 Hardware 
Error 
H Errors occurred to: 
1. External A/D sampling 2.5 V 
2. 12V power supply voltage 
3. Internal A/D sampling 2.5 V 
4. Pump. 
5. 3-way valve. 
CO2 Sampleline 
Occluded 
M An error or occlusion occurred to the sampling line. 
CO2 Zero Failed  H Deviation of gain input signal is too great to be adjusted. 
Namely, it cannot be adjusted within the normal range: 3.5 
V±100 mV. 
CO2 Cal. Failed  M The difference between the measured standard gas 
concentration and the specified standard gas concentration 
exceeded 40% of the specified standard gas concentration. 
Or, an illegal calibration parameter was obtained. The 
normal parameter calibration range is within 0.2 to 2.5. 
CO2 System Error  H Multiple system errors occurred. 
CO2 No Watertrap  M The CO
2 
watertrap fell off or was not connected. 
EtCO2 Overrange  H
FiCO2 Overrange  H
The measured value is outside of the measurement range. 
Contact your service personnel.   
CO2 Check Airway    M Airway error. 
CO2 No 
Sampleline 
L Make sure if the sampling line is already connected. 
CO2 Main Board 
Error 
H
CO2 Check Sensor 
or Main Board 
M
CO2 Replace 
Scrubber&Pump 
M
CO2 Replace 
Sensor 
M
CO2 15V 
Overrange 
M
CO2 module malfunction. Re-plug the module or restart 
the anesthesia machine. 
BIS Init Error  H
BIS Selftest Error  H
BIS Comm Stop  H
BIS Comm Error  H
Module malfunction or communication failure between 
the module and main unit. Re-plug the module, restart the 
machine, or plug the module onto other main unit. 
BIS 
module 
BIS ALM LMT 
ERR 
M The alarm limits are changes accidentally. Contact your 
service personnel.