D-13
BIS Overrange H
SQI Overrange H
SR Overrange H
The measured value is outside of the measurement range.
Contact your service personnel.
BIS High Imped. M
BIS Sensor Off L
Check sensor connections. Re-connect the sensor.
BIS DSC Error M BIS DSC signal reception error. Check the DSC.
BIS DSC Malf H BIS DSC switched off due to malfunction. Check the
DSC.
BIS No Cable L Connect the BIS cable.
BIS No Sensor L Connect the sensor.
BIS Wrong Sensor
Type
L Check or replace the sensor.
SQI<50% L
SQI<15% L
SQI is too low. Check the patient’s condition and sensor
connection.
BIS Sensor Expired L Replace the sensor.
BIS Sensor Failure M Put the sensor again or replace the sensor.
BIS Sensor Too
Many Uses
L Replace the sensor.
Disconnect/Reconn
ect BIS
H Re-plug the BIS module.