DMTA-10022-01EN, Rev. C, January 2015
Custom Setups for Single Element Transducers
169
IF BLANK in modes 2 and 3 only (see “Interface Blank” on page 180 for
details)
ECHO 2 DETECT in modes 2 and 3 only (see “Detection of Echo 1 and
Echo 2” on page 179 for details)
ECHO 1 DETECT (see “Detection of Echo 1 and Echo 2” on page 179 for
details)
ECHO WINDOW (see “Detection of Echo 1 and Echo 2” on page 179 for
details)
MB BLANK (see “Main Bang Blank” on page 175 for details)
TDG SLOPE (see “TDG Slope” on page 175 for details)
INIT GAIN (see “Initial Gain” on page 175 for details)
MAX GAIN (see “Maximum Gain” on page 174 for details)
PULSER POWER (see “Pulser Power” on page 172 for details)
PROBE TYPE
MEAS TYPE (see step 5.c in “Creating a Custom Setup for a Single Element
Transducer” on page 165 for details)
4. Use the [] or [] keys to select the value for the selected parameter.
5. Repeat steps 3 and 4 to adjust other parameters.
6. Press [GAIN/WAVE ADJ] again to hide the waveform adjustment parameter.
10.3 Detection Modes
There are three detection modes (Mode 1, Mode 2, and Mode 3):
Mode 1
Uses a direct contact transducer to measure the time of flight between the main
bang and the first back-wall echo (see Figure 10-3 on page 170).