EasyManua.ls Logo

Olympus 45MG - Figure 12-2 the KEYPAD TEST Screen

Olympus 45MG
246 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10022-01EN, Rev. C, January 2015
Chapter 12
204
3. In the TESTS screen, select the desired test, and then press [ENTER].
4. When KEYPAD TEST is selected, in the KEYPAD TEST screen (see Figure 12-2
on page 204), perform the following steps:
a) Test any key by pressing it on the keypad.
When that specific key is working correctly, the instrument displays the name
of the key in the Last Key pressed box.
b) Press [ENTER] to terminate the keypad test.
Figure 12-2 The KEYPAD TEST screen
5. When VIDEO TEST is selected:
a) In the VIDEO TEST screen, press [MEAS] to start the video test.
The display shows vertical color strips. Any problems with the display show
up as a disruption of the pattern.
b) Press [ENTER] to terminate the video test.
6. When INTERNAL SD CARD TEST or EXTERNAL SD CARD TEST is selected,
in the INTERNAL SD CARD TEST or EXTERNAL SD CARD TEST screen:
a) Press [MEAS] to start the test.
The test result appears under SD Card Test Status. The possible results are as
follows:
PASS: Indicates that the card is functioning properly.
FAIL: Indicates that there is a problem with the memory card. When the
external card fails, reinstall or replace the card, and then restart the
instrument. When the internal card fails, contact Olympus for service.
b) Press [ENTER] to terminate the SD CARD test.

Table of Contents

Other manuals for Olympus 45MG