EasyManua.ls Logo

Olympus 45MG - 12.10 Resolving Measurement Problems; Table 16 Measurement Troubleshooting

Olympus 45MG
246 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10022-01EN, Rev. C, January 2015
Chapter 12
208
12.10 Resolving Measurement Problems
Table 16 Measurement troubleshooting
Symptom Possible explanation
No echoes or weak echoes,
and no measurement (LOS)
There is insufficient couplant, especially on
rough or curved surfaces.
The gain is set too low.
The material is too attenuating, or the surfaces
are unparalleled or too rough.
The gage requires service; try the master reset.
The echo is too low in amplitude to be
detected. Try increasing the gain.
Strong echoes, no
measurements
The echo may be in a blank area of the
waveform, and cannot be detected.
The echo is outside of the echo window.
Strong echoes, incorrect
measurements
Out of calibration; perform calibration.
Differential mode — check DIFF flag.
Min or Max mode — see “Using the
Minimum, Maximum, or Min/Max Thickness
Mode” on page 129.
The material is too grainy, has flaws,
inclusions or laminations, or very high surface
noise; try manual gain adjustment or
extended blank.

Table of Contents

Other manuals for Olympus 45MG

Related product manuals