DMTA-10022-01EN, Rev. C, January 2015
Chapter 10
170
Figure 10-3 Mode 1 detection example
Mode 2
Measures the time of flight between the interface (or delay line) echo and first
back-wall echo using a delay line or an immersion transducer (see Figure 10-4 on
page 170).
Figure 10-4 Mode 2 detection example
Mode 3
Uses a delay line or an immersion transducer to measure the time of flight
between one back-wall echo and the next back-wall echo (see Figure 10-5 on
page 171).
Main bang
First back-wall echo
Detection marker
Mode indicator
Main bang
First back-wall echo
Interface echo
Detection bar
Mode indicator