EasyManua.ls Logo

Olympus 45MG - Dual Element Transducer Selection in Echo-To-Echo Modes; Figure 6-5 Comparing Manual Measurements

Olympus 45MG
246 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10022-01EN, Rev. C, January 2015
Chapter 6
74
a) In the measurement screen, press [SETUP], and then select MEAS.
b) In the MEAS menu, set the MEAS OPTION to MANUAL E-TO-E, and then
press [MEAS].
2. Press [GAIN/WAVE ADJ].
The wave adjustment parameter appears (see Figure 6-5 on page 74).
Figure 6-5 Comparing manual measurements
3. Use the [] and [] keys to select either the EXT BLANK or the E1 BLANK
parameter.
4. Use the [] and [] keys to adjust the value to exclude unwanted peaks and to
detect the desired echoes.
6.2.2 Dual Element Transducer Selection in Echo-to-Echo Modes
Although the optional Echo-to-Echo modes work with all the 45MG dual element
transducers, Olympus recommends using specific transducers for particular thickness
ranges in steel parts (see Table 4 on page 75).
Poor manual measurement Good manual measurement
[WAVE ADJ] parameter

Table of Contents

Other manuals for Olympus 45MG

Related product manuals