DMTA-10022-01EN, Rev. C, January 2015
Appendix A
212
Measurement resolution Selectable from the keypad:
LOW: 0.1 mm (0.01 in.)
STD: 0.01 mm (0.001 in.)
HI: 0.001 mm (0.0001 in.) with the High Resolution option.
Not all resolutions are available for all measurement modes.
Material sound velocity range 0.508 mm/µs to 18.699 mm/µs (0.0200 in./µs to 0.7362 in./µs)
Material sound velocity
resolution
0.001 mm/µs (0.0001 in./µs)
Alarm set point range 0.00 mm to 500.00 mm (0.00 in. to 20.00 in.)
Table 24 Datalogger specifications
Parameter Value
Storage capacity 475000 thickness readings or
20000 waveforms with thickness readings
ID number length 1 to 20 characters
File name length 1 to 32 characters
File formats Incremental
Sequential (defined by starting and ending ID number)
Sequential with custom points
2-D grid
Boiler
External memory card microSD memory card
2 GB maximum capacity
Table 25 Typical ranges and default setups for single element
transducers
a
Setup name Transducer Typical measurement range
DEFM1-20.0-M116 M116 Steel: 0.250 mm to 8.000 mm (0.020 in. to 1.500 in.)
DEFM1-10.0-M112 M112 Steel: 0.760 mm to 250.000 mm (0.030 in. to 10.000 in.)
DEFM1-10.0-M1016 M1016 Steel: 0.760 mm to 250.00 mm (0.030 in. to 10.000 in.)
DEFM1-5.0-M110 M110 Steel: 1.00 mm to 380.00 mm (0.040 in. to 15.000 in.)
DEFM1-5.0-M109 M109 Steel: 1.00 mm to 500.00 mm (0.050 in. to 20.000 in.)
DEFM1-2.25-M106 M106 Steel: 2.00 mm to 635.00 mm (0.080 in. to 25.000 in.)
Table 23 Measurement specifications (continued)
Parameter Value