EasyManua.ls Logo

Philips HeartStart XL+ - Page 22

Philips HeartStart XL+
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2: Maintenance Software Upgrades
12
7 Use the Up/Down buttons to select the desired software release and language.
8 Press the Select button to begin the upgrade.
Press the Main Service soft key to exit the upgrade screen. Do not attempt to cancel the upgrade
after it has started.
CAUTION: Power interruption during software installation may disable the device and require an extensive repair.
After the upgrade starts, be careful not to remove the USB drive or power source, or turn the Therapy
Knob in order to not interrupt the software installation process.
The software is installed on the device. This process takes a few minutes. While the software is being
updated, progress messages are displayed and the soft keys are disabled, see Figure 9 on page 13.
NOTE: The screen may change appearance or blank out when the new software is loaded to the Processor PCA,
the RFU may display a red X, and chirps and alarms may sound—all these are normal behavior.
Do not interfere with the upgrade process!
9 Depending on the Software Releases currently on your device and the one being installed, the
installation may proceed in several stages; in this case you may see the “Stage x of y” text.
CAUTION: Remember NOT to remove the USB drive or power source, or turn the Therapy Knob!
Disregard any technical alarms that may be displayed in the process.
Figure 8 Software Upgrade, Sample Selection Screen








