Operational Check 3: Troubleshooting
37
Leads ECG Test: Depending on your configuration and attached cables, the test may display:
Pass/ECG Cable
•Pass/No Cable
Fail/ECG Cable
Fail/No Cable
Fail
If a possible reason of the test failure might be a faulty cable connection, then the following prompt
is displayed at the end of all remaining tests:
Leads ECG Test Failed With Cable.
Disconnect ECG Cable to Rerun Test Without Cable.
Disconnect the ECG cable or acknowledge the Proceed Without Rerunning Test prompt.
Pads/Paddles ECG Test: Depending on your configuration and attached cables, the test may display:
Pass/Pads
Pass/External Paddles
•Pass/No Cable
Fail/Pads
Fail/External Paddles
Fail/Internal Paddles
Fail/No Cable
Battery Test: The test may display:
•Pass
Fail
None (if battery is not inserted)
•Calibration Required
Low Battery (charge the battery; calibration is not required)
SpO
2
Test: Performed only if the SpO
2
option is installed. The test may display:
•Pass
Fail
EtCO
2
Tes t: Performed only if the EtCO
2
option is installed. Depending on the sensor model, the test
may display:
•Pass
Fail
No Sensor Detected
•Calibration Overdue
Replace Sensor
NBP Test: Performed only if the NBP option is installed. The test may display:
•Pass
Fail
•Calibration Overdue
Printer Test: Among other actions, Printer tests determines if the correct font is installed in the printer. If
the Printer Font is not installed in the printer, but is available in the device memory, then the Printer
Font is installed in the printer, while the message Printer Font Download in Progress is displayed.
Depending on the AutoPrint configuration setting, a successful Printer Test displays either the
Check Printed Report or Print Op Check Report to Check Printer prompt.