5 Patient Alarms and INOPs Technical Alarm Messages (INOPs)
68
CO
2
CHANGE SCALE CO
2
The CO
2
wave is clipped. Select a more appropriate wave scale
to display the whole wave.
CO
2
CHK ADAPTER
Numeric is replaced by a -?-
INOP tone.
CO
2
Check that the sensor is connected to the airway adapter, clean
the airway adapter, if necessary. Perform a zero calibration. If
the INOP persists, contact your service personnel.
CO
2
CHECK CAL
Numeric is replaced by a -?-
INOP tone.
CO
2
The CO
2
value is outside the measurement range. Perform an
accuracy check for both calstick cells and, if necessary,
recalibrate the transducer.
CO
2
DEACTIVATED
INOP tone.
CO
2
The CO
2
measurement label in the measurement device has
been deactivated by deactivating the label in the Measurement
Selection window. The measurement automatically disappears
from the display. To switch the measurement on again,
reactivate the measurement label in the Measurement Selection
window.
C.O. DEACTIVATED
INOP tone.
C.O. The Cardiac Output measurement label in the measurement
device has been deactivated by deactivating the label in the
Measurement Selection window. The measurement
automatically disappears from the display. To switch the
measurement on again, reactivate the measurement label in the
Measurement Selection window.
C.O. EQUIP MALF
Numeric is replaced by a -?-
INOP tone.
C.O. There is a problem with the C.O. hardware. Contact your
service personnel.
CO
2
EQUIP MALF
Numeric is replaced by -?-
INOP tone.
CO
2
The Measurement Extension is faulty. Unplug and replug the
Multi-Measurement Module with Extension. If you are using
the mainstream method, unplug and replug the transducer or
try another transducer. If the INOP persists, contact your
service personnel.
CO
2
FAILED CAL
Numeric is replaced by -?-
INOP tone.
CO
2
Make sure that the Cal cell was changed between CAL1 and
CAL2. Repeat the calibration. If the INOP reappears, try
another transducer. If the INOP persists, contact your service
personnel.
CO
2
NO SENSOR
Numeric is replaced by -?-
INOP tone.
CO
2
There is no CO
2
sensor connected. If you silence this INOP
the CO
2
measurement will be switched off.
CO
2
NO TRANSDUC
Numeric is replaced by -?-
INOP tone.
CO
2
There is no CO
2
transducer connected. If you replace the
transducer, the new transducer must be calibrated. If you
silence this INOP the CO
2
measurement will be switched off.
CO
2
NO TUBING
Numeric is replaced by -?-
INOP tone.
CO
2
Either the sample line is disconnected, or an incorrect line is
attached. Check the connection. If necessary, connect another
sample line (Use only the approved accessories).
If you silence this INOP, the measurement will be switched off.
(!!/!!!)CO
2
OCCLUSION
Numeric is replaced by a -?-
INOP tone.
CO
2
The sample line or exhaust tube is blocked. Check the tubing,
then disconnect and reconnect the sample line. If the INOP
persists, connect a new sample line.
CO
2
OVERRANGE
Numeric is replaced by -?-
INOP tone.
CO
2
The CO
2
value is higher than the measurement range. If you
suspect a false high value, contact your service personnel.
CO
2
PUMP OFF
Numeric is replaced by a -?-.
CO
2
The pump has been switched off for fifteen minutes. To switch
it on again, select Pump On in the Setup CO
2
menu.
INOP Message, Indication Source What to do