Technical Alarm Messages (INOPs) 5 Patient Alarms and INOPs
69
CO
2
PURGING
Numeric is replaced by a -?-
INOP tone.
CO
2
The Filterline is being purged to remove an occlusion in the
line or airway adapter. If the occlusion is removed, the INOP
will disappear. If not, the INOP
CO
2
OCCLUSION is
displayed.
CO
2
SENS. WARMUP
Numeric is displayed with a -?-
Microstream CO
2
: INOP tone.
Mainstream CO
2
: no INOP tone
CO
2
Wait until the sensor reaches operating temperature and the
INOP disappears.
C.O. UNPLUGGED
numeric is replaced by -?-
INOP tone.
C.O. Plug in the C.O. module. Silencing this INOP switches off the
measurement.
CO
2
UPDATE FW
Numeric is replaced by a -?-
INOP tone.
CO
2
The software in the Measurement Extension does not match
the software in the MMS. Contact your service personnel.
CO
2
WAIT CAL2
Numeric is replaced by a -?-
CO
2
Calibration on the first calstick cell is complete. Place the
transducer on the other calstick cell and start the CAL2
calibration cycle.
CO
2
ZERO FAILED
Numeric is replaced by a -?-
INOP tone.
CO
2
An error occurred during the last zero calibration. Check the
airway adapter and clean, if necessary. Perform another zero
calibration. If the INOP persists, contact your service
personnel.
CO
2
ZERO REQU’D
Numeric is replaced by a -?-
INOP tone
CO
2
Perform zero calibration for the CO
2
sensor. If the INOP
persists, contact your service personnel.
CO
2
ZERO RUNNING CO
2
Wait until zero calibration is finished.
CPP CHK SOURCES
Numeric is replaced by a -?-
CPP Not all measurements or values required to perform the
calculation are available. Check the measurement sources.
CPP CHK UNITS
Numeric is replaced by a -?-
CPP The monitor has detected a conflict in the units used for this
calculation. Check the unit settings.
!!/!!!CUFF NOT DEFLAT
Numeric is displayed with a -?-
Severe yellow/red INOP tone.
During this INOP, alarms cannot be
paused or switched off.
NBP Remove the cuff from the patient. Make sure that the tubing is
not kinked or twisted and that the correct patient category is
selected. Try repeating the measurement.
You can silence the INOP, but the INOP message remains
visible until the next NBP measurement is started or the
Stop All SmartKey is selected.
[Adult or pediatric patients: The NBP cuff pressure has exceeded
15mmHg (2kPa) for more than 3 minutes.
Neonatal patients: The NBP cuff pressure has exceeded 5mmHg
(0.7kPa) for more than 90 seconds.]
!!/!!!CUFF OVERPRESS
Numeric displayed with -?- ;
Severe yellow/red INOP tone.
During this INOP, alarms cannot be
paused or switched off.
NBP The NBP cuff pressure exceeds the overpressure safety limits.
Remove the cuff from the patient. Make sure that the tubing is
not kinked or twisted and that the correct patient category is
selected. Try restarting the measurement.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the
Stop
All
SmartKey is selected.
CVP INOPS PRESS See <Pressure label> INOPS (under Pressure).
DEVICE CHECK SETUP
INOP tone.
IntelliBridge Device identification completed, but communication could
not be established due to timeout.
IntelliBridge INOP abbreviations may differ slightly depending
on the device category.