Chapter 3 Calibration
Revision W HF Series X-ray Generators - Service Manual
3-36 Quantum Medical Imaging, LLC
4. On the voltmeter, observe the measured voltage value. Ideally, the value
appearing in the Value in Gen (top right field) will match the reading on
the voltmeter. If not, adjust the value in the Measure #1 field up or
down by pressing the key next to the up or down arrow, respectively,
until the Measure #1 field is set to the value displayed on the voltmeter
(to within Calibration Accuracy ranges shown in Table 3-5). Each time
the key is pressed changes the voltage value up/down in one-tenth volt
increments.
5. Press key next to the Save Calibrate field to save the calibration entry.
6. Press the key next to Previous Screen to return to the A/D Calibration
menu.