Chapter 3 Calibration
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging, LLC 3-55
* Only required on 150 kVp systems
Table 3-8. AEC Calibration Worksheet
Film/Screen
Combination:
Table 40 Wall 40 Wall 72 Other1 Other2
AEC OD CALIBRATION
Setting:
50 kVp, 200L mA, 5 cm
Optical Density:
AEC Ref.
RECIPROCITY
Setting:
First Cal. - 50 kVp, 400L mA, 15 cm
Optical Density:
Setting:
Middle Cal. - 50 kVp, 75S mA, 15 cm
Optical Density:
Setting:
Save Cal. - 50 kVp, 25S mA, 15 cm
Optical Density:
AEC KV CALIBRATION
Setting:
Step 1 - 50 kVp, 200L mA, 5 cm
Optical Density:
Setting:
Step 2 - 60 kVp, 200L mA, 10 cm
Optical Density:
Setting:
Step 3 - 80 kVp, 200L mA, 15 cm
Optical Density:
Setting:
Step 4 - 100 kVp, 200L mA, 20 cm
Optical Density:
Setting:
Step 5 - 110 kVp, 200L mA, 25 cm
Optical Density:
Setting:
Step 6 - 120 kVp, 200L mA, 30 cm
Optical Density:
Setting:
*Step 7 - 130 kVp, 200L mA, 30 cm
Optical Density:
Setting:
*Step 8 - 140 kVp, 200L mA, 30 cm
Optical Density: