EasyManuals Logo

Quantum ODYSSEY HF Series User Manual

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #147 background imageLoading...
Page #147 background image
Chapter 3 Calibration
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging, LLC 3-55
* Only required on 150 kVp systems
Table 3-8. AEC Calibration Worksheet
Film/Screen
Combination:
Table 40 Wall 40 Wall 72 Other1 Other2
AEC OD CALIBRATION
Setting:
50 kVp, 200L mA, 5 cm
Optical Density:
AEC Ref.
RECIPROCITY
Setting:
First Cal. - 50 kVp, 400L mA, 15 cm
Optical Density:
Setting:
Middle Cal. - 50 kVp, 75S mA, 15 cm
Optical Density:
Setting:
Save Cal. - 50 kVp, 25S mA, 15 cm
Optical Density:
AEC KV CALIBRATION
Setting:
Step 1 - 50 kVp, 200L mA, 5 cm
Optical Density:
Setting:
Step 2 - 60 kVp, 200L mA, 10 cm
Optical Density:
Setting:
Step 3 - 80 kVp, 200L mA, 15 cm
Optical Density:
Setting:
Step 4 - 100 kVp, 200L mA, 20 cm
Optical Density:
Setting:
Step 5 - 110 kVp, 200L mA, 25 cm
Optical Density:
Setting:
Step 6 - 120 kVp, 200L mA, 30 cm
Optical Density:
Setting:
*Step 7 - 130 kVp, 200L mA, 30 cm
Optical Density:
Setting:
*Step 8 - 140 kVp, 200L mA, 30 cm
Optical Density:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals