Chapter 5 Service Instructions
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging, LLC
5-9
Anode
Volt
(Cont’d)
3. If waveform is correct, go to step 4.
If not, replace KVP Control Board
A2 and re-check drive signal as
described in the previous step. If
problem persists, replace Logic
Board A1 (AY40-006S) and re-
check. If still not corrected, check
wiring.
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action
APPROX.
1.2 MS
DELAY

Table of Contents