EasyManua.ls Logo

Quantum ODYSSEY HF Series

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 5 Service Instructions
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging, LLC
5-11
Anode Volt
(Cont’d)
6. Verify AC line voltage meets speci-
fication during an exposure: Check
capacitor voltage (C9-C12 and
C13-C16 on single-phase units,
C1-C4 on SE units, or C9-C12 and
C13-C16 on three-phase units) is
between +350 and +440 VDC.
The power must be sufficient to
support the exposure (i.e., check
for weak AC Line).
7. Verify that all grounds are good
(JP2-JP9 installed or shorted on
Logic Board A1 (AY40-006S)).
APR Tree
Load Error
{text
string}
An error occurred while loading
the APR tree. Note: {text string}
will contain additional details.
Call Technical Service/Support.
Backup Backup time was reached based
on current technique factors.
1. Reduce technique factor(s). Re-
take exposure.
2. Verify ion chamber cable connec-
tions are correct and secure.
3. On AEC Control Board A11
(AY40-031S or AY40-027S), check
+15 VDC and -15 VDC outputs are
not loaded down. If so, remove ion
chamber cable connection and re-
check voltages. If +15 VDC and /or
-15 VDC outputs are now good, ion
chamber is at fault. If still loaded
down, replace AEC Control Board.
4. Verify the Backup Time/mAs set-
tings in APR view (if caused during
APR mode) and/or Backup Time/
mAs in Service Settings configura-
tion screen are set correctly (i.e.,
not too low).
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action

Table of Contents

Related product manuals