Chapter 5 Service Instructions
Revision W HF Series X-ray Generators - Service Manual
5-24 Quantum Medical Imaging, LLC
KVP
Volts
Indicates the exposure kVp volt-
age is out of range.
Call Technical Service/Support.
LBMast
Micro
Logic Board master microproces-
sor A1U9 returned incorrect
checksum.
Check A1U9. Call Technical Service/
Support.
LBSlave
Micro
Logic Board slave microproces-
sor A1U10 returned incorrect
checksum.
Check A1U10. Call Technical Ser-
vice/Support.
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action